Figure 10.  Verify patient and sensor information.






















               Step 2: Acquire Signal
               Once you have loaded the patient and sensor information into the Hospital Electronics System and then implanted the sensor, the
               steps below describe how to set the pulmonary artery (PA) pressure baseline.
               1.   Place the antenna under the patient’s back, approaching from the side where the sensor was placed. Antenna is not sterile.
                   Sliding the antenna between the sheet and the table works well. Position the center of the antenna under the tip of the patient’s
                   shoulder blade on the side of the torso where the sensor is located.
               2.   The signal strength (SS) bar indicates the robustness with which the Hospital Electronics System is measuring the implanted
                   sensor. If the signal strength bar is optimal (>70% and a green color, see figure below), go to the next step. If not, go to the
                   section Optimize Signal Strength.

               Step 3: Set Baseline
               1.   A pulmonary artery or Swan-Ganz  catheter is used to set the baseline pressure for the sensor. For an accurate sensor
                                          1
                   measurement, it is important to set up the pulmonary artery or Swan-Ganz catheter properly. See Pulmonary Artery Catheter
                   Setup section.
               2.   Initially, you will observe that the pressure values on the right of the waveform are dashed out. This indicates that the PA
                   pressure baseline has not been set.

               Figure 11.  Initial Pressure Waveform.






















               3.   To Set the PA Pressure Baseline, wait for the screen to fill with a stable pressure waveform and then select the Set Pressure
                   Baseline button.
               4.   Enter the mean pressure from the pulmonary artery catheter measurement and select the OK button.





               1  Swanz-Ganz is a registered trademark of Edwards Lifesciences Corporation.